In this example, the instrument is set up to log samples only if the temperature of the sample is
above 18

°

C.

To alter the switch so that data is logged when the specific conductance is greater than 100 uS/cm
and temperature is greater than 18

°

C, first press 5 for SpCond mS/cm.

The prompt 0-lower or 1-higher than (1) will appear on the screen under the menu.

First type 1 (higher than) and press Enter.

The screen prompt will ask for the condition: Threshold (0.000):

Enter your new condition in uS/cm (100) and press Enter.

The screen will now display the altered menu which shows the new condition and data will be
logged only when the temperature is greater than 18

°

C AND the specific conductance is greater

than 100 uS/cm.

========================================
Set-Condition
----------------------------------------
1-(*) Temp C

F-( ) DO %

2-( ) Temp F

G-( ) DO mg/L

3-( ) Temp K

H-( ) Depth ft

4-( ) SpCond mS/cm

I-( ) Depth m

5-( ) SpCond uS/cm

J-( ) pH

6-( ) Cond mS/cm

K-( ) pH mV

7-( ) Cond uS/cm

L-( ) ORP mV

8-( ) Resist MOhm*cm

M-( ) NH4+ mg-N/L

9-( ) Resist KOhm*cm

N-( ) NH3 mg-N/L

A-( ) Resist Ohm*cm

O-( ) NO3- mg-N/L

B-( ) TDS g/L

P-( ) Turb NTU

C-( ) TDS kg/L

Q-( ) TDG mm-Hg

D-( ) TDS mg/L

R-( ) Batt V

E-( ) Sal ppt

Condition: Temp C > 18.000

Select option (0 for previous menu):