YSI 6000UPG User Manual

Page 154

Advertising
background image

E-3

(6) After writing the DO and pH compensation files to the C:\PC6000 directory, type "PC6000" at
the "C:\PC6000" prompt to re-enter the PC6000 software package. Just as would be done for no
compensation, prepare the plot and report output format desired by manipulation of the
"Parameters", the "Control Information", and the "Scaling Information".

(7) After the format is finalized, re-enter the "Parameters" section of the software. Place the cursor
on the "DO (mg/l)" entry, hit "Return", and then select the "Detail this parameter" option within the
window.

(8) Move the cursor to the "Compensate" entry and hit "Return". Move the cursor to the "DO,
mg/l" entry and hit "Return" to activate the compensation of this parameter.

(9) Hit the "Tab" key once to move the cursor to the right portion of the screen and then enter the
location and designation of the DO compensation file in this bar. In this case, the entry is:
C:\PC6000\SRUN7DO.CMP. Hit "Return" and the screen should appear as follows if the
procedure has been followed:

(10) Carry out the same procedure for pH:

(a) Activation of the compensation, and
(b) Entry of the location and designation of the pH compensation file.

* NOTE: Although the probe part number does not correspond to any of the 6000

UPG

probes, the

probe identification of 6030 is correct for this application.

Parameter:

DO (mg/L)

Formula:

Temp (C) Cond (mS/cm) DO (%)

Inputs:

Temperature (C) from 6030* probe, channel 0
Conductivity (mS/cm) from 6030 probe, channel 1
DO (%) from 6030 probe, channel 2

Compensate:

DO (mg/L)

c:\pc6000\srundo.cmp