Ysi 6026 turbidity probe – YSI 6000UPG User Manual

Page 21

Advertising
background image

2-11

5.

Install the 6083 or 6084 probe into the sonde as described

below.

A. Apply a very thin coat of O-ring lubricant (supplied in the
YSI 6040 Maintenance Kit) to the O-ring on the connector side
of the probe and to the O-ring on the probe housing. Insert the
probe assembly into the selected sonde port.

B. Screw the probe assembly into the sonde. Use the hex driver
to assist you in tightening the probe into the sonde.
DO NOT OVER TIGHTEN.

1

. Remove the turbidity probe from its box and any protective caps from the probe.

2.

Locate the plug which seals the auxiliary (AUX) port

on the sonde. Using the hex driver to assist you, remove
the plug.

3.

Install the 6026 probe into the sonde as described

below.

A.

Apply a very thin coat of O-ring lubricant (supplied in

the 6040 Maintenance Kit) to the O-ring on the
connector side of the probe. Insert the probe in the port
marked AUX.

B

. With the connectors aligned, screw down the probe

nut, making sure that the probe and sonde connectors are

YSI 6026 Turbidity Probe

Hex Driver

O-ring