3 sdi-12 interface – YSI 6000UPG User Manual

Page 130

Advertising
background image

9-2

9.3 SDI-12 INTERFACE

SDI-12 is an industry standard serial digital interface bus. The bus was designed to allow
compatibility between data collection devices and sensors of various manufacturers. The
description below applies specifically to the 6000

UPG

implementation of SDI-12 interface. For

complete SDI-12 technical specifications please contact:

Campbell Scientific, Inc.
P.O. Box 551
Logan, Utah 84321
USA
(801) 753-2342

SDI-12 is a single master multi-drop bus. As many as 10 sensors can be connected to the bus at a
time. Each sensor is preassigned a unique address from 0 to 9. Each 6000

UPG

is factory set to

address 0. The address can be changed in the System Setup menu (see Section 4.5). The 6000

UPG

implements the basic SDI-12 command set. Below are the descriptions of each command and their
responses. At the end of this section, you will find the 6000

UPG

commands that are used to run and

troubleshoot SDI-12.

The following notations are used:

a

6000

UPG

SDI-12 address (ASCII '0' to '9')

[CR]

Carriage return (ASCII 13)

[LF]

Line feed (ASCII 10)

Master

Any SDI-12 compatible data collection device

Signal

Dir

Function

COLOR

-----------------------------------------------------------------------------------------------------------------------

Shield

Shield ground

BARE

SDIDAT

I/O

SDI-12 data

PURPLE

TX

OUT

RS-232 Transmit data

YELLOW

SGND

-

RS-232 signal ground

GREEN

RX

IN

RS-232 Receive data

ORANGE

PWR

IN

SDI-12 power

RED

PGND

-

SDI-12 ground

BLACK