2-7

5.

Locate the plug which seals the DO.C.T. port on the

sonde. Using the hex driver to assist you, remove the
plug.

6.

Install the 6030 probe into the sonde as described

below.

A. Apply a very thin coat of O-ring lubricant (supplied
in the 6040 Maintenance Kit) to the O-ring on the
connector side of the probe. Insert the probe in the port
marked DO.C.T.

B. With the connectors aligned, screw down the probe
nut, making sure that the probe and sonde connectors
are fully meshed. Use the hex driver to assist you in
tightening the nut. DO NOT OVER TIGHTEN.

7.

Replace the probe guard and tighten the probe guard

mounting screws.

O-ring