F-1

APPENDIX I

SYSTEM ERRORS AND WARNINGS

WARNING: Low RTC battery

Lithium battery that powers the real time clock is low.
Replace lithium battery.

ERROR(000): Undetermined

Internal error. The system is unable to determine the cause.
Repeat action. If necessary reset by removing the battery lid
and repeat action.

ERROR(001): Illegal input conversion

Internal error. The system encountered an unknown
character in the last I/O conversion. Repeat action.

ERROR(002): Conversion overflow

Input number exceeds system limit.

ERROR(003): Input error

User input error.

ERROR(004): Input out of range

Input number exceeds allowable range.

ERROR(005): Illegal hex nibble

System encountered bad data during system software update.

ERROR(006): Illegal address

System encountered bad address during system software
update.

ERROR(007): Illegal record

System encountered bad record during system software
update.

ERROR(008): Illegal page number

System encountered bad code page during system software
update.

ERROR(009): Mismatch hex checksum

System has detected bad checksum during system software
update.

ERROR(010): Hex load abort

System has detected abort sequence during system software
update.

ERROR(011): SFLS erase suspend

Internal error. System software erasure was suspended.

ERROR(012): SFLS erase fail

Fatal hardware failure. Unable to remove old code during
system software update. Battery is low or program failed to
save system setup information.

ERROR(013): SFLS program fail

Fatal hardware failure. Failed to update system software or
failed to save system setup.

ERROR(014): SFLS VPP low

Fatal hardware failure. Unable to raise programming voltage
or low battery voltage.

ERROR(015): SFLS command sequence

Internal error. Bad programming sequence during system
software update.

ERROR(016): Write to occupied sflscell

Internal error. Hardware failure during system software
update.

ERROR(017): SC timeout

Internal error.

ERROR(018): SQ empty

Internal error.

ERROR(019): SQ full

Internal error.

ERROR(020): Rx overrun

Internal error.

ERROR(021): Rx framming

Internal error.

ERROR(022): Rx parity

Internal error.

WARNING: High DO charge

DO charge current exceeds limit (25 to 75 during calibration.
Probe maintenance is required.

WARNING: Out of range

Calibration procedures determine that actual solution and
user inputted number cannot be resolved.

ERROR: Sensor disabled

System is not configured with appropriate sensor for
calibration.