YSI 6000UPG User Manual

Page 18

Advertising
background image

2-8

STEP 2. INSTALL OTHER PROBES

If you have purchased any optional YSI probes, follow instructions provided below for all items
other than the 6029 conductivity/temperature sensor which is installed in an identical fashion to the
6030. The 6031 pH, 6032 ORP, 6033 low ionic strength pH, 6084 nitrate, and 6083 ammonium
probes can be ordered as options and do not require factory installation (see Section 4.7 for
software setup). The 6026 turbidity probe is installed by the user according to its accompanying
instructions, but requires an additional PCB which must be factory installed. The depth sensors are
factory installed options.

Follow these instructions to prepare your new probe and install it in the 6000

UPG

.

NOTE:

Instructions for preparation of use, maintenance and storage are included with each

electrode.

1.

Remove the storage bottle or hydrating cap and rinse the

glass portion of the electrode with distilled water.

2.

Apply a very thin coat of O-ring lubricant (supplied in the

YSI 6040 Maintenance Kit) to the O-ring on the glass shaft of
the new electrode. Insert the electrode into the probe housing
and hand tighten.

DO NOT OVER TIGHTEN.

YSI 6031 pH Probe
YSI 6032 ORP Probe
YSI 6033 Low Ionic Strength pH Probe

Hydrating Cap
or Storage

l

Probe