7 advanced setup – YSI 6000UPG User Manual

Page 96

Advertising
background image

5-24

you study. While the discrete sample mode is still operative, exit. to the PC6000 menu bar by
pressing
F10 (do not press Esc or the sampling will be terminated).

STEP 6: From the PC6000 menu bar, select Real-Time and press Enter. The Discrete Sample
data from the sonde will be plotted on the display and will be logged to disk automatically under
your designated file name.

COMMON CAUSES OF ERRORS IN REAL-TIME MODE:

1. The “Parser” is not set to “6000”. Set the Parser to “6000”

2. You have selected more parameters in the Sonde Report Setup than can be viewed on the

computer screen. Remove time and date and/or other less important parameters in the Report
Setup Menu so that all selected parameters are visible in Discrete Sample display.

3. The name which you entered for your data file already exists on your storage disk, but the file

format for the initial study is different from the current file format. Change the name of the
data file or delete the existing data file of this name from your disk.

4. You exited the Discrete Sample mode of the sonde software using Esc rather than F10.

Return to the Sonde menu and select Run. Restart the Start sampling mode and exit with
F10 to the PC6000 menu bar. Choose Real-Time and press Enter.

5.7 ADVANCED SETUP

From the PC6000 menu bar, select Setup. The setup menu will be displayed.

+------------------------------------------------------------------------------+
¦

Sonde

File

Setup

Real-Time

Exit

YSI PC6000

¦

¦------------------------------------------------------------------------------¦
¦

¦

¦

¦

¦

¦

¦

-----DISPLAY-FORMATS-----

-----------COMPUTER-CONNECTIONS-----------

¦

¦

Date format:

mdy

Comm port:

2

¦

¦

Date separator:

/

Baud rate:

9600

¦

¦

Time separator:

:

Printer port:

1

¦

¦

Radix mark:

.

Printer type:

HP LaserJet laser

¦

¦

¦

¦

¦

¦

-------MORE-SETUP--------

-------------COMPUTER-DISPLAY-------------

¦

¦

Data capture setup...

Menu colors:

Bering Sea (color)

¦

¦

Advanced Setup...

Plot colors:

Ohio River (color)

¦

¦

Graphics mode:

Automatic selection hi-res

¦

¦

¦

¦

¦

¦

¦

¦

¦

¦

¦

¦------------------------------------------------------------------------------¦
¦

Press Enter to access advanced setup items

¦

+------------------------------------------------------------------------------+