2 calibrate, Conductivity – YSI 6000UPG User Manual

Page 58

Advertising
background image

4-9

This message will disappear when the unattended sampling is over and you will be able to interact
with the sonde via the menu. You will be then be given the option to discontinue the unattended
sample as shown below.

To stop the Unattended study, enter “Y” and press Enter.

4.2 CALIBRATE

All of the 6000

UPG

sensors (except temperature) require periodic calibration to assure high

performance. Select 2. Calibrate from the sonde Main menu, the Calibrate menu will be
displayed.

Select this option to calibrate the conductivity probe and a second menu will offer you the options
of calibrating in Specific Conductance, Salinity, or Conductivity. After selecting the option of
choice (Specific Conductance is normally recommended), you will be asked to enter the value of
the standard used during calibration. Be certain that the units are correct. After pressing Enter,

** INSTRUMENT IS IN UNATTENDED MODE **

Start date and time

: 05/31/96 11:04:08

Duration (days)

: 30.000

Interval (minutes)

: 2

Site description

: First-test

UNATTENDED mode active. Cancel (Y/N)?

========================================
Calibrate
----------------------------------------
1. Conductivity

6. pH

2. DO%

7. Ammonium

3. DO mg/L

8. Nitrate

4. Depth

9. Turbidity

5. ORP

A. TDG

Select option (0 for previous menu):

1.

Conductivity