YSI 6000UPG User Manual
Page 127
Advertising
8-4
SENSOR PERFORMANCE PROBLEMS continued...
“High DO Charge”
message during DO cal
DO sensor needs reconditioned
Recondition sensor. Use 6035 Kit
7.2
Tubidity probe unstable
or inaccurate. Error
messages appear during
calibration
Probe requires cleaning.
Follow probe cleaning procedure
7.2
Probe requires calibration
Follow cal procedures
4.2
Probe has been damaged
Replace probe
E
Water in probe connector
Dry connector; reinstall probe
E
Calibration solutions out of spec
Use new calibration solutions
Wiper is not turning or is not
synchronized.
Activate wiper. Assure rotation.
Make sure set screw is tight.
7.4
Wiper is fouled or damaged.
Clean or replace wiper.
7.4
Internal failure.
Return sonde for service.
C