SENSOR PERFORMANCE PROBLEMS continued...

“High DO Charge”
message during DO cal

DO sensor needs reconditioned

Recondition sensor. Use 6035 Kit

7.2

Tubidity probe unstable
or inaccurate. Error
messages appear during
calibration

Probe requires cleaning.

Follow probe cleaning procedure

7.2

Probe requires calibration

Follow cal procedures

4.2

Probe has been damaged

Replace probe

E

Water in probe connector

Dry connector; reinstall probe

E

Calibration solutions out of spec

Use new calibration solutions

Wiper is not turning or is not
synchronized.

Activate wiper. Assure rotation.
Make sure set screw is tight.

7.4

Wiper is fouled or damaged.

Clean or replace wiper.

7.4

Internal failure.

Return sonde for service.