YSI 6000UPG User Manual

Page 134

Advertising
background image

9-6

6000

UPG

:

1+071592+101500+17.5+12.05+98.7[CR][LF]

Master:

1D1!

6000

UPG

:

1+8.25+6.45-325+10[CR][LF]

After finishing the measurement, the 6000

UPG

sent a service request to indicate completion. The

bus master then sent the "D0" command to retrieve the data. There were 5 data returned. Since 9
readings should be available, the master continued with "D1" commands and received the
remaining data.

Master:

1D1!

6000

UPG

:

1+8.25+6.45-325+10

For some reason, the bus master received scrambled data with the previous "D1" command. It
reissued the command to get the correct data. The responses from "D0" and "D1" commands are

Date:

07/15/92

Time:

10:15:00

Temperature:

17.5

Specific conductance:

2.05

DO

%:

98.7

DO

mg/L:

8.25

pH

(ISE1):

6.45

ORP

(ISE2):

-325

Depth:

10

RUNNING SDI-12:

Four commands are available to run and troubleshoot SDI-12. These commands are entered from
the command line, at the '#' prompt. Following are the descriptions of each command.

SDIADDR [a]

Set the 6000

UPG

SDI-12 address to 'a'. The address 'a' is from 0 to 9. If 'a' is greater than 9, the

address remains unchanged. If 'a' is not specified, the 6000

UPG

will display the current SDI-12

address.