F-2

Error Table continued...

ERROR: Need 2 at same conc

A 3 point ammonium or nitrate calibration requires that two
of the three calibration solutions be of the same
concentration, but of significantly different temperature.

ERROR: Illegal value

During calibration an illegal value has been entered.

ERROR: Cal canceled by user

User has canceled calibration.

ERROR(038): FDDT allocation

Internal error.

ERROR(039): Unformatted Flash disk

Flash disk has never been formatted or has been corrupted.
Format the flash disk.

ERROR(040): File not found

Incorrect file number.

ERROR(041): Invalid file number

Incorrect file number.

ERROR(042): Unable to create new file

Number of files in Flash disk has reached 23. The Flash disk
can only hold a maximum of 23 files including all erased
files. Upload all existing files then format the Flash disk.

ERROR(043): Flash disk full

Flash disk is full. Upload all existing files then format the
Flash disk.

ERROR(044): BAD Flash disk. Check Vpp

Fatal hardware failure. Programming voltage failed. Check
battery voltage.

ERROR(045): End-of-file

Internal error.

ERROR(046): Bad Flash disk block

Internal error.

ERROR(047): Invalid block number

Internal error.

ERROR(048): Unexpected end-of-file

Internal error.

ERROR(049): Bad file link

Internal error.

ERROR(050): Unknown trap service

Internal error.

ERROR(051): Invalid record number

Internal error.

ERROR(052): Bad record

Internal error.

ERROR(053): Unable to format Flash disk

Flash disk is degrading. System failed to format Flash disk.
Reformat Flash disk.

ERROR(054): Flash disk test fail

Fatal hardware failure; if Flash disk format was successful
but Flash disk test failed.

ERROR(055): File not found

Internal error.

ERROR(056): No active file

Tried to close an unopened file.

ERROR(057): Kermit upload canceled

Kermit file transfer has detected an aborted sequence.

ERROR(058): Kermit upload timeout

Kermit file transfer timeout. Communication link is broken
or the host system failed to initiate a Kermit receiving
sequence.

ERROR(059): Unattended mode active

Tried to run a conditional sample while the system is in
unattended mode.

ERROR(060): File open

When the Model 6000 is in unattended mode or the 1st data
point is logged to flash disk from either a discrete or
unconditional sample, the system time and date and sensor
setup are not accessible.

ERROR(061): Discrete sample not active

Internal error.

ERROR(062): Stop time less than start time

Internal error.

ERROR(063): Stop time less than current time

Internal error.

ERROR(064): Unknown

Internal error.