Appendix e application note – YSI 6000UPG User Manual

Page 152

Advertising
background image

E-1

APPENDIX E

APPLICATION NOTE

SENSOR DRIFT COMPENSATION WITH THE 6000

UPG

The following application note is designed to facilitate the use of the sensor drift software included
in the PC6000 package by providing the user with a specific example. The use of this function is
generalized in Section 5.4 of the 6000

UPG

manual.

Situation

After laboratory calibration, a user deployed their 6000

UPG

in a readily-accessible stream on

10/14/92 with the instrument set to log data to memory at 15 minute intervals. The study was named
SRUN7. The instrument was left in place for approximately 2 weeks and during this time period,
the user visited the site 3 times, took a water sample in a sealed bottle, and, on returning to the
laboratory, measured the dissolved oxygen and pH with recently-calibrated instruments. The user
also collected a water sample when the instrument was retrieved (10/26/92) and measured dissolved
oxygen and pH in the laboratory. The following records were kept.

Laboratory Measurements

Date

Time

DO (mg/L)

pH

10/17/92

8:00

5.40

7.62

10/29/92

8:00

7.50

7.63

10/23/92

10:00

7.30

7.58

10/26/92

8:15

6.32

7.60

The user now has everything required to compensate the DO and pH readings for field drift:

The data file logged to sonde memory.

The correct DO and pH readings at various times during the study based on grab samples of

water which were analyzed with recently-calibrated laboratory instruments.

PC6000 software for retrieval and manipulation of the data from the sonde memory.