8-3

SENSOR PERFORMANCE PROBLEMS continued...

SYMPTOM

POSSIBLE CAUSE

ACTION

REF

pH, ORP, Ammonium,
or Nitrate unstable or
inaccurate. Error
messages appear during
calibration.

Probe requires cleaning

Follow probe cleaning procedure

7.2

Probe requires calibration

Follow cal procedures

3.1

Probe has been damaged

Replace pH or ORP probe.
Replace sensor module on
ammonium or nitrate sensors.

E

Water in probe connector

Dry connector; reinstall probe

E

Calibration solutions out of spec

Use new calibration solutions

ORP fails Zobell check

Repeat test. Check possible
causes above.

6.3

Internal failure

Return sonde for service

C

Depth unstable or
inaccurate

Depth sensor has not been zeroed

Follow depth zero procedure

3.1

Depth sensor access hole is
obstructed

Follow depth cleaning procedure

7.2

Depth sensor has been damaged

Return sonde for service

C

Internal failure

Return sonde for service

C

Conductivity unstable or
inaccurate. Error
messages appear during
calibration.

Conductivity improperly calibrated

Follow cal procedure

3.1

Conductivity probe requires
cleaning

Follow cleaning procedure

7.2

Water in probe connector

Dry connector; reinstall probe

2.2

Conductivity probe damaged

Replace 6030 probe

2.2

Calibration solution out of spec

Use new calibration solution

3.2

Calibration value too far from
sample value

Cal at or near expected value

3.2

Internal failure

Return sonde for service

C

Calibration solution or sample
does not cover entire sensor.

Immerse sensor fully.

3.1

Installed probe has no
reading

The sensor has been disabled

Enable sensor

4.7

Water in probe connector

Dry connector; reinstall probe

2.2

Probe has been damaged

Replace the probe

2.2

Report output improperly set up

Set up report output

4.6

Sonde’s autodetect feature has
malfunctioned

Return sonde for service

C

Internal failure

Return sonde for service.

C