Ion concentration

ProLab 4000

104

ba75646e02

01/2010

ISE Cal

This is the conventional two-point to nine-point calibration proce-
dure
that uses 2 to 9 freely selectable standard solutions. The concen-
tration expected in the measurement determines the concentration of
the calibration standards.

Stability control

During calibration, the stability control is automatically activated.

The current measurement with stability control can be terminated at
any time (accepting the current value).

Calibration record

The new calibration values are displayed when a calibration procedure
is finished. Then you can decide whether you want to take over these
values of the new calibration or whether you want to continue measur-
ing with the old calibration data. After accepting the new calibration val-
ues the calibration record is displayed.

Display of calibration

data and download to

interface

You can display the data of the last calibration (see page 160). Subse-
quently, you can transmit the displayed calibration data to the interface,
e. g. to a printer or PC, with the <PRINT> key.

Sample record:

Note
The calibration record is automatically transmitted to the interface
after calibrating.

Print date

26.04.07 16:13

ProLab 4000 (06249876)
Current user: 1234567890

CALIBRATIONISE
Calibration date

20.04.07 10:14:03

User: Administrator

Standard 1:

0.010 mg/l

Standard 2:

0.020 mg/l

Voltage 1:

0.0 mV 24.0 °C

Voltage 2:

9.0 mV 24.0 °C

Ion type:

Ag

Slope:

29.9 mV

Sensor

+++