6 system functions, 1 overview: system settings, System functions – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 56: Overview: system settings, 6system functions

Advertising
background image

System functions

ProLab 4000

56

ba75646e02

01/2010

6

System functions

System functions are all functions that are independent of the probe.

6.1

Overview: System settings

The following sensor-independent meter features can be adjusted in
the System menu and its submenus:

Menu item

Setting

Description

System

Data output...

USB printer

( )

USB interface /Baud

( ) / 1200 ... 115200

Baud rate of the data interface

Serial interface /Baud

( ) / 1200 ... 115200

Baud rate of the data interface

Output format

ASCII
CSV

Output format for data transmis-
sion
For details, see page 179

Autom. printout...

Printout interval

Automatic output of the measure-
ment data to the interfaces at the
specified interval.

Reset

-

Resets the system settings to the
default values.
For details, see page 66

Measuring

Autom. stability control

( )

You can activate or switch off the
automatic stability control func-
tion (see page 63).

Temperature unit

Celsius
Fahrenheit

All temperatures are displayed
with the selected unit.

Connect Ref1/Ref2

( )

Use reference electrode of the
other pH/ISE input