Ion concentration

ProLab 4000

130

ba75646e02

01/2010

11

In the Standard register, enter the values for Standard volume
and Standard concentration if necessary.

12

Add the standard solution to the sample.

13

Start the measurement with [Continue].
The measured value is checked for stability (Stability control).

14

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.

15

Change the displayed unit of the measured value with [Unit] if
necessary.

16

If required, display the record of the current measurement with
[Record].

17

Start measuring further samples with [Continue].

Repeat the steps 5 - 15 for all samples.
or
Terminate the measuring method with [Cancel] and confirm the
safety query with [OK].
Measurement with the selected method is completed.