ProLab 4000

Conductivity

137

ba75646d02

01/2010

Temperature compensa-

tion for diluted solutions

For the measurement of aqueous solutions

z in a low concentration range (diluted solution)

z in a small temperature range (|T - Tref| < 10 K)

z with a fixed reference temperature at 20 °C or 25 °C,

the procedures of linear (TC lin) and nonlinear (TC nLF) temperature
compensation are suitable.

The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
[TR20] or [TR25].

Application tips

Select the following temperature compensation given in the table
according to the respective test sample:

Note
The reference temperature and temperature compensation are set in
the Measuring menu (see page 134).

Test sample

Temperature compensation

Display
indica-
tor

Natural water

(ground water, surface
water, drinking water)

nLF

according to EN 27 888

nLF

Ultrapure water

nLF

according to EN 27 888

nLF

Seawater
(salinity)

Automatic nLF according to IOT

Sal, nLF

Other diluted solutions
with known tempera-
ture dependency

Linear temperature compensa-
tion;

Temperature coefficient can be
set in the range

0.001 ... 3.000 %/K

lin

Solutions with
unknown temperature
dependency

No temperature compensation