Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 108

Advertising
background image

Ion concentration

ProLab 4000

108

ba75646e02

01/2010

Continuing with two-

point calibration

9

Wait for the measurement with stability control to end.
or
Take over the calibration value with [OK].

The ISE standard input window for the entry of the concentra-
tion of the second standard solution pops up.

10

Select a value from the ISE-Std list.
or
Enter a value and, if necessary, an exponent with <0...9>.
or
Cancel the calibration with [Cancel].

Note
To enter a concentration in exponential representation such as

1.00 E2 mg/l:

Enter the mantissa with <0...9> (here: 1.00).

Enter the exponential sign (E) with <- .><- .>.

Enter the exponent with <0...9> (here: 2).

11

When measuring without temperature sensor:
z Measure the temperature of the standard solution using a

thermometer.

z Open the setting window for the temperature value with

[Man. temp.].

z Enter the temperature and confirm with [OK].

12

Thoroughly rinse the electrode with distilled water.

13

Immerse the electrode in standard solution 2.

14

Start the measurement with [OK].
The measured value is checked for stability (Stability control).

15

Wait for the measurement with stability control to end.
or
Take over the calibration value with [OK].
The ISEX calibration window for the next standard solution
pops up and the Finish with 2-point calibration infotext is dis-
played.