16 what to do if, 1 ph and orp measurement, Chapter 16 w – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 189: To chapter 16 w

Advertising
background image

ProLab 4000

What to do if...

189

ba75646e02

01/2010

16

What to do if...

16.1

pH and ORP measurement

Error message Err1

Error message Err2

Error message Err4

Error message, Error

Cause

Remedy

pH electrode:

– Measured value outside the

measuring range

– Use a suitable electrode

– Air bubble in front of the dia-

phragm

– Remove air bubble

– Air in the diaphragm

– Extract air or moisten junc-

tion

– Cable broken

– Replace the electrode

– Gel electrolyte dried out

– Replace the electrode

Cause

Remedy

– No electrode connected

– Connect the electrode

– Setting time during calibration

too long

– Adjust temperature if neces-

sary

– Recalibrate

Cause

Remedy

– Temperature not stable during

calibration.

– Adjust temperature if neces-

sary

– Recalibrate

Cause

Remedy

pH electrode:

– The values determined for zero

point and slope of the electrode
are outside the allowed limits.

– Recalibrate

– Diaphragm contaminated

– Clean diaphragm