ProLab 4000

What to do if...

191

ba75646e02

01/2010

CalClock on red

background

16.2

Ion selective measurement

Error message Err1

Obviously incorrect

measured values

Error message, Error

(invalid calibration)

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Measuring range exceeded

– Dilute test sample

Cause

Remedy

– Electrode not connected

– Connect the electrode

– Cable broken

– Replace the electrode

Cause

Remedy

ISE electrode:

– Moisture in the plug

– Dry plug

– Electrode obsolete

– Replace the electrode

– Electrode unsuitable for the

range to be measured

– Use a suitable electrode

– Socket damp

– Dry socket

Calibration procedure:

– Wrong sequence of standards

for three point calibration

– Select correct sequence

– Calibration standards do not

have the correct temperature
(max.

±

2 °C temperature differ-

ence)

– Adjust the temperature of

the calibration standards