ORP voltage

91

ba75646e02

01/2010

8.2.2

Measuring

Stability control

TheStability control function checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of
measured values. The background color is red until the measured
value is stable.

Regardless of the setting for Autom. stability control (see page 63) in
the System menu, you can start the Stability control function manually
at any time.

Note
ORP electrodes are not calibrated. However, you can check ORP
electrodes using a test solution.

1

Perform the preparatory activities (see page 69).

2

Highlight the channel of the electrode in the measured value
display.

3

If necessary, select the U display with <MODE>.

4

Immerse the ORP electrode in the test sample.

1

Highlight the channel of the ORP electrode.

2

If necessary, select the U or dU parameter with <MODE>.

3

Freeze the measured value with <SC>.
The display switches to the display of the selected channel as
necessary.

In the status line, [SC] is displayed.