Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 190

Advertising
background image

What to do if...

ProLab 4000

190

ba75646e02

01/2010

No stable measured

value

Obviously incorrect

measured values

– Electrode broken

– Replace the electrode

Buffer solutions:

– Incorrect buffer solutions

– Select the suitable buffer set

for the buffers used

or
– Replace the buffer solutions

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Replace the buffer solutions

Cause

Remedy

Cause

Remedy

pH electrode:

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

Test sample:

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Adjust temperature if neces-

sary

Electrode + test sample:

– Conductivity too low

– Use a suitable electrode

– Temperature too high

– Use a suitable electrode

– Organic liquids

– Use a suitable electrode

Cause

Remedy

pH electrode:

– Not connected

– Connect the electrode

– Cable broken

– Replace the electrode

– pH electrode unsuitable

– Use a suitable electrode

– Temperature difference

between buffer and test sam-
ple too high

– Adjust temperature of buffer

or sample solutions