10 selecting and determining the nonlinear, Temperature compensation tc nli – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 146

Advertising
background image

Conductivity

ProLab 4000

146

ba75646d02

01/2010

10.2.10 Selecting and determining the nonlinear temperature

compensation TC nLin4

Preparatory activities

The nonlinear temperature compensation TC nLin4 is determined from
several standards of unknown concentration.

You prepare the required standards yourself through equidistant dilu-
tion of the standard with the highest concentration e.g.:

Standard with the high-

est concentration

Volume share in %

Deionized water

Volume share in %

Standard 1

100

0

Standard 2

95

5

Standard 3

90

10

Standard 4

85

15

Standard 5

80

20

1

Highlight the channel of the conductivity measuring cell in the
measured value display.

2

If necessary, select the measured parameter with <MODE>.

3

Immerse the measuring cell in the test sample.

4

Open the input window for the measurement of the tempera-
ture dependency with Calibration / TC calibration / TC nLin3.

Advertising