Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 82

Advertising
background image

pH

ProLab 4000

82

ba75646e02

01/2010

Continuing with three-

point to five-point

calibration

(Buffer ) DIN)

Note
A calibration line is determined for the two-point calibration.

16

When measuring without temperature sensor:
z Measure the temperature of the standard solution using a

thermometer.

z Open the setting window for the temperature value with

[Man. temp.].

z Enter the temperature and confirm with [OK].

17

Thoroughly rinse the electrode with distilled water.

18

Immerse the electrode in buffer solution 3.

19

Start measurement of the next buffer with [Continue].
The measured value is checked for stability (Stability control).

20

Wait for the measurement with stability control to end.
or
Terminate the stability control with [OK].
The pHX calibration window for the according to buffer solution
pops up and the Finish with 3-point calibration infotext is dis-
played.

21

Using [OK], confirm the Finish with 3-point calibration infotext
and take over the new calibration values.
The calibration record is displayed and output to the interface.
The calibration procedure is finished.
or

Continue the calibration with [Continue].
or
Cancel the calibration procedure without taking over the cali-
bration values with [Cancel], or <ESC>.

Note
The calibration is automatically finished after all buffers of a buffer set
were used.

Note
The calibration line is determined by linear regression.