Ion concentration

ProLab 4000

118

ba75646e02

01/2010

10

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Standard register opens up.

11

Add standard solution to the sample (2 percent by volume of
the sample solution).

12

Start the measurement with [Continue].
The measured value is checked for stability (Stability control).

13

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.

14

Change the displayed unit of the measured value with [Unit] if
necessary.

15

If required, display the record of the current measurement with
[Record].

16

Start measuring further samples with [Continue].

Repeat the steps 2 - 14 for all samples.
or
Terminate the measuring method with [Cancel] and confirm the
safety query with [OK].
Measurement with the selected method is completed.