6.8

Automatic stability control

The Autom. stability control function continuously checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values.

You can activate or switch off the Autom. stability control function
(menu, Autom. stability controlSystem).

The Autom. stability control function is carried out:

z as soon as the measured value is outside the allowed stability range

z when you switch over between the measured parameters with

<MODE>.

The stability of the measured value is indicated by the background
color of the measured value. In the delivery condition, the SC stable /
SC not stable conditions are assigned to the following colors:

Color

Meaning

red

SC not stable
Stability criteria not met

green

SC stable
Stability criteria met

SC stable

SC not stable