Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 129

Advertising
background image

ProLab 4000

Ion concentration

129

ba75646e02

01/2010

Measuring

6

In the Sample register, enter the following values:
z the volume of the sample (Sample volume)
z the volume of the ISA/TISAB solution (ISA/TISAB volume)
z the volume of the blank value solution (Blank value volume)
z the concentration of the blank value solution (Blank value

conc.).

7

Thoroughly rinse the electrode with deionized water.

8

Immerse the electrode in the sample that was supplemented
with blank value solution.

9

Start the measurement with [Continue] .

The measured value is checked for stability (Stability control).

10

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Standard register opens up.