Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 150

Advertising
background image

Conductivity

ProLab 4000

150

ba75646d02

01/2010

Sample record:

Calibration evaluation

After the calibration, the meter automatically evaluates the current sta-
tus of the calibration. The evaluation appears on the display and in the
calibration record.

Print date

26.04.07 16:13

ProLab 4000 (06249876)
Current user: 1234567890

CALIBRATION Cond
Calibration date

20.04.07 10:14:03

LF413-ID (A062498765)
User: Administrator

Calibration interval:

150 days

Standard:

0,01 mol/l KCl

Cond./Tref 25:

1413 µS/cm

Cal. temp:

25,4 °C

Cell constant:

0.475 1/cm 25.0 °C

Sensor

+++

Note
If non-ID sensors are used, the sensor type and series number are not
output in the calibration record.

Display

Calibration record

Cell constant [cm

-1

]

+++

within the ranges

0.450 ... 0.500 cm

-1

0.585 ... 0.715 cm

-1

or

0.800 ... 1.200 cm

-1

Error

Error

outside the ranges

0.450 ... 0.500 cm

-1

or

0.800 ... 1.200 cm

-1

Eliminate the error according to chapter
16 W

HAT

TO

DO

IF

... (page 189)