Calibration points

Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The measuring instrument
determines the following values and calculates the calibration line as
follows:

19

Reagecon TEC *

2,000
4,010
7,000

10,000

25 °C

20

Reagecon 20 *

2,000
4,000
7,000

10,000
13,000

20°C

21

Reagecon 25 *

2,000
4,000
7,000

10,000
13,000

25 °C

22

Riedel-de Haen *

2,000
4,000
7,000

10,000

20°C

No.

Buffer set *

pH values

at

Note
The buffers are selected in the Calibration sensor menu (see
page 73).

Displayed calibration data

1-point

z Zero point
z Slope = Nernst slope (-59.16 mV/pH at 25 °C)

2-point

z Zero point
z Slope

3-point to
5-point

z Zero point
z Slope

The calibration line is calculated by linear regression.