01/2010

9.4

Blank value correction

If the ion concentration in the sample is so small that it is not in the lin-
ear range of the electrode, you can increase the ion concentration in
the linear range with the blank value correction by the addition of a
blank value solution.

The displayed measured value is the difference between the actual
measured value and the determined blank value.

Calibration

Measuring

1

In the Measuring / ISE Setup menu, select and confirm the
menu item, Blank value correction.
The Blank value correction function is active (

).

2

Carry out a 2 to 5-point calibration according to the user guide
(see page 106).
As soon as a stable value is achieved for the standard calibra-
tion solution, the slope (mV) and evaluation (Error, +++) of the
electrode after calibration is displayed.
The infotext, Accept calibration and terminate is displayed.

3

Using [OK], confirm the infotext Accept calibration and termi-
nate
.
The calibration record is displayed and output to the interface.
The calibration procedure is finished.

4

Close the calibration record with [OK].
The prompt for the measurement of the blank value appears.

5

Thoroughly rinse the electrode with deionized water.

6

Immerse the electrode in the blank value solution.

7

Start the measurement with [Continue].
The measured value is checked for stability (Stability control).

8

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The calibration record is displayed and output to the interface.
The calibration procedure is finished.

9

Close the calibration record with [OK].