6 measuring with increment procedure (methods), 1 selecting the measuring method, Measuring with increment procedure (methods) 9.6.1 – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 113: Selecting the measuring method

Advertising
background image

ProLab 4000

Ion concentration

113

ba75646e02

01/2010

9.6

Measuring with increment procedure (methods)

9.6.1

Selecting the measuring method

The following methods are supported:

z Standard addition
z Double standard addition
z Standard subtraction
z Sample addition
z Standard subtraction
z Blank value addition (Standard addition with blank value correction)

1

Perform the preparatory activities (see page 103).

2

Highlight the channel of the electrode in the measured value
display.

3

If necessary, select the ISE display with <MODE>.

4

Thoroughly rinse the electrode with distilled water.

5

Adjust the temperature of the calibration standard solutions.

6

Select a unit in the menu, Measuring / ISE settings / Units .

7

Select a method in the menu, Measuring / ISE methods .
The selected method is started (see page 114 ff).

Advertising