1 carrying out calibration, Carrying out calibration – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 106

Advertising
background image

Ion concentration

ProLab 4000

106

ba75646e02

01/2010

9.3.1

Carrying out calibration

Use two to nine standard solutions in ascending or descending order.

Preparatory activities

Perform the following preparatory activities when you want to calibrate:

Calibration

Note
When calibrating within the framework of a method measurement, it is
always a two-point calibration.

1

Connect an ion sensitive electrode to the meter.

2

Highlight the channel of the electrode in the measured value
display.

3

If necessary, select the ISE display with <MODE>.

4

If necessary, open the Measuring / ISE settings / Ion type menu
and select the ion type to be measured.

5

Keep the standard solutions ready.

6

If necessary, measure the temperature of the standard solu-
tions with a thermometer.

7

If necessary, change the unit of the measurement result and
calibration standards in the menu, Measuring/Units.

1

Perform the preparatory activities (see page 106).

2

Start the calibration with <CAL>.

The ISEX calibration window opens.