2 measuring, Measuring – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 99

Advertising
background image

ProLab 4000

Ion concentration

99

ba75646e02

01/2010

9.2.2

Measuring

Temperature for

calibrating and

measuring

For precise ion selective measurements, the temperature difference
between measurement and calibration should not be greater that 2 K.
Therefore, adjust the temperature of the standard and measuring solu-
tions accordingly. If the temperature difference is greater the [TempErr]
warning appears in the measured value display.

Stability control

The Stability control function continuously checks the stability of the
measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The background color is red until
the measured value is stable.

Regardless of the setting for automatic Autom. stability control (see
page 63) in the System menu, you can start the Stability control func-
tion manually at any time.

1

Perform the preparatory activities according to page 97.

2

Immerse the electrode in the test sample.

1

Highlight the channel of the electrode in the measured value
display.

2

If necessary, switch to the ISE display with <MODE>.

3

Freeze the measured value with <SC>.
In the status line, [SC] is displayed.

Advertising