Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 123

Advertising
background image

ProLab 4000

Ion concentration

123

ba75646e02

01/2010

6

In the Standard register, enter the values for Standard volume-
Standard concentration and ISA/TISAB volume if necessary.

7

Thoroughly rinse the electrode with deionized water.

8

Immerse the electrode in the standard.

9

Start the measurement with [Continue].

The measured value is checked for stability (Stability control).

10

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Sample register opens up.

11

Add the sample to the standard solution.

12

In the Sample register, enter the values for the sample volume
(Sample volume) if necessary.

13

Start the measurement with [Continue].
The measured value is checked for stability (Stability control).

14

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.