2 standard addition – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 114

Advertising
background image

Ion concentration

ProLab 4000

114

ba75646e02

01/2010

9.6.2

Standard addition

In the Standard addition procedure, a known amount of standard solu-
tion is added to the sample.

The ion concentration in the sample is calculated from the change in
potential.

Calibration

Measuring

1

Check the currently displayed ion.
If the ion to be measured is displayed incorrectly:

Select the ion type in the menu, Measuring / ISE methods / Ion
type
.

2

In the Measuring / ISE methods menu, select the method,
Standard addition.

The ISE standard input window for the first calibration standard
appears.

3

Carry out a two-point calibration according to the user guid-
ance (see page 106).
As soon as a stable value is achieved for the second standard
calibration solution, the slope (mV) and evaluation (Error, +++)
of the electrode after calibration is displayed.
The infotext, Accept calibration and terminate is displayed.

4

Using [OK], confirm the infotext Accept calibration and termi-
nate
.
The calibration record is displayed and output to the interface.
The calibration procedure is finished.

5

Close the calibration record with [OK].
The ISE method: The Standard addition window pops up. The
Sample register is displayed.