1 determining the cell constant (calibration) – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 151

Advertising
background image

ProLab 4000

Conductivity

151

ba75646d02

01/2010

10.3.1

Determining the cell constant (calibration)

Proceed as follows to determine the cell constant:

1

Connect a conductivity measuring cell to the meter.

2

Select the channel of the measuring cell in the measured value
display.

3

If necessary, select the parameter, conductivity with <MODE>.

4

Start the calibration with <CAL>.

The calibration display appears.

5

Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.

6

Start the measurement with [OK].
The measured value is checked for stability (Stability control).

7

Wait for the end of the measurement with Stability control or
accept the calibration value with [OK].
The cell constant determined is displayed.

8

Accept the new calibration values with [OK].

The calibration record is displayed and output to the interface.
The calibration procedure is finished.
or
To switch to the measured value display without accepting the
new calibration values, press <ESC>.