3 conductivity measurement – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 192

Advertising
background image

What to do if...

ProLab 4000

192

ba75646e02

01/2010

Warning [TempErr]

Warning [ISEErr]

16.3

Conductivity measurement

Conductivity measuring

cell was not recognized

Error message, Error

(invalid calibration)

CalClock

with red frame

Cond temp?

Cause

Remedy

– Temperature difference

between measurement and cali-
bration greater than 2 K.

– Adjust the temperature of

the test sample

Cause

Remedy

– Electrode voltage outside cali-

brated range

– Recalibrate

Cause

Remedy

– Measuring cell not connected

– Connect measuring cell

– Cable broken

– Replace measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Unsuitable control standard

– Use control standard,

0.01 mol/l KCl

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Test sample does not have the

correct temperature (max.

±

2

°C temperature difference from
calibration temperature)

– Adjust the temperature of the

test sample