3 calibration settings and calibration data, Calibration settings and calibration data – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 87

Advertising
background image

ProLab 4000

pH

87

ba75646e02

01/2010

7.3.3

Calibration settings and calibration data

In the Calibration menu, you can

z set the calibration interval,

z define the calibration type and

z view the calibration data.

Menu item

Setting

Description

Calibration

pH calibration type

AutoCal

( )

Calibration with commercial buffer sets.

The buffer set is selected with the next
menu item (Buffer sets)

Buffer sets

DIN ( )
TEC
...

Buffer sets to be used for pH calibration.
For details, see page 74.

VariCal

( )

Calibration with any buffer solutions

Slope in

mV/pH ( )
%

Unit of the slope.
The % display refers to the Nernst slope of
-59.16 mV/pH ([determined slope/Nernst
slope ] x 100).

Zero point in

mV ( )
pH

Unit for the zero point.

Calibration interval...

1 ... 999

Calibration interval... for the pH electrode
(in days).
The meter reminds you to calibrate regu-
larly by the red frame around the CalClock
in the measuring window.

Calibration history...

Displays the calibration history of the last
calibrations.

Advertising