Conductivity

ProLab 4000

140

ba75646d02

01/2010

10.2.7

Selecting and adjusting the nonlinear
temperature compensation TC nLin1

The nonlinear temperature compensation is carried out according to
the following equation:

The equation includes two temperature coefficients, TC

1

and TC

2

. In

the ProLab 4000 meter, the temperature coefficients TC

1

and TC

2

are

stored for the following test samples:

Suitable temperature coefficients TC

1

and TC

2

for solutions that are not

mentioned here can be taken from tables found in further reading and
then entered in the meter.

X

Tref

= X

T

/ (1 + TC

1

·[T-TRef] + TC

2

·[T-TRef]

2

)

X

T

= Conductivity at temperature T

[S/cm]

X

Tref

= Conductivity at reference temperature Tref

[S/cm]

TC

1

= Temperature

coefficient

1

[1/°C]

TC

2

= Temperature

coefficient

2

[1/°C

2

]

T

= Temperature

[°C]

Tref

= Reference temperature

[°C]

HCl

NaOH

NaCl

KCl

Concentration c
[mol/l]

1,0

0,67

0,85

0,67

Temperature range
[°C]

0 … 55

0 … 40

0 … 40

0 … 40

Tref [°C]

25

25

25

25

1

Highlight the channel of the conductivity measuring cell in the
measured value display.

2

If necessary, select the measured parameter with <MODE>.

3

Open the input window for the temperature coefficients with
Calibration / TC calibration / TC nLin1....