5 sample addition – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 122

Advertising
background image

Ion concentration

ProLab 4000

122

ba75646e02

01/2010

9.6.5

Sample addition

In the Sample addition procedure, a known amount of test sample is
added to the standard solution.

The ion concentration in the sample is calculated from the change in
potential.

Calibration

Measuring

1

Check the currently displayed ion.
If the ion to be measured is displayed incorrectly:

Select the ion type in the menu, Measuring / ISE methods / Ion
type
.

2

In the Measuring / ISE methods menu, select the method,
Sample addition.

3

Carry out a two-point calibration according to the user guid-
ance (see page 106).
As soon as a stable value is achieved for the second standard
calibration solution, the slope (mV) and evaluation (Error, +++)
of the electrode after calibration is displayed.
The infotext, Accept calibration and terminate is displayed.

4

Using [OK], confirm the infotext Accept calibration and termi-
nate
.
The calibration record is displayed and output to the interface.
The calibration procedure is finished.

5

Close the calibration record with [OK].
The ISE method: The Sample addition window pops up. The
Standard register is displayed.