Technical Data

ProLab 4000

200

ba75646e02

01/2010

Cell constants

Cell constant C

Values

Can be calibrated in
the ranges

0.450 ... 0.500 cm

-1

0.585 ... 0.715
0.800 ... 1.200 cm

-1

Adjustable

0.010 cm

-1

(fixed)

0.090 ... 0.110 cm

-1

0.250 ... 25.000 cm

-1

Reference temperature

Reference tempera-
ture

Values

Adjustable

20 °C (Tr20)
25 °C (Tr25)

Accuracy (± 1 digit)

Variable

Accuracy

Temperature of the test
sample

ϰ

/ Temperature compensation

None (Off)

± 0.5 %

Nonlinear (nLF)

± 0.5 %

0 °C ... + 35 °C
according to EN 27 888

± 0.5 %

+ 35 °C ... + 50 °C
Extended nLF function

according to SI Analytics
GmbH measurements

Linear (lin)

± 0.5 %

+ 10 °C ... + 75 °C

SAL / range

0.0 ... 42.0

± 0.1

+ 5 °C ... + 25 °C

± 0.2

+ 25 °C ... + 30 °C

TDS [mg/l]

± 1