Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 107

Advertising
background image

ProLab 4000

Ion concentration

107

ba75646e02

01/2010

3

When measuring without temperature sensor:
z Measure the temperature of the standard solution using a

thermometer.

z Open the setting window for the temperature value with

[Man. temp.].

z Enter the temperature and confirm with [OK].

4

Thoroughly rinse the electrode with distilled water.

5

Immerse the electrode in standard solution 1.

6

Open the ISE standard input window with [Continue].

7

Select a value from the Concentration list.
or
Enter a value with <0...9>.

or
Cancel the calibration with [Cancel].

Note
To enter a concentration in exponential representation such as

1.00 E2 mg/l:

Enter the mantissa with <0...9> (here: 1.00).

Enter the exponential sign (E) with <- .><- .>.

Enter the exponent with <0...9> (here: 2).

8

Start the measurement with [OK].
The measured value is checked for stability (Stability control).