Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual

Page 145

Advertising
background image

ProLab 4000

Conductivity

145

ba75646d02

01/2010

7

Warm or cool the measuring solution (standard 1) to the start
temperature Start temp..
As soon as the temperature of the measuring point is achieved
and a stable value is recognized, the measurement is automat-
ically started and the dataset stored.
After the measurement, a piece of information about the next
measuring point with the relevant nominal temperature is dis-
played and the temperature difference between the current
temperature and the nominal temperature.

8

Slowly warm or cool the measuring solution (standard 1) until
the end temperature End temp. is achieved.
Each time the nominal temperature for a measuring point is
reached a dataset is stored.
At the end of the series of measurements for the solution (stan-
dard 1), the corresponding measuring record is displayed.

9

Confirm the record with [Continue].

10

Repeat the steps 5 to 8 for all further solutions (standard 2 to x).

The corresponding calibration record with the determined tem-
perature coefficients is displayed at the end of the calibration.
If the calibration is successful, the determined temperature
dependency for the TC nLin3 method is stored and used for fur-
ther measurements with this method.

11

Confirm the record with [Continue].
[nLin3] is displayed in the status line of the measuring channel.