CONTREX CX-1200 User Manual

Page 264

Advertising
background image

7 - 25

Data Read Constant Table

7) Data Read Parameter Title
8) Data Read Maximum Parameter Value
9) Data Read Minimum Parameter Value
10) Data Read Default Parameter Value
11) Data Read Backup Parameter Value
12) Data Read Value of the Maximum Parameter Field length
13) Data Read Parameter Control Byte

Note:

For all constant table reads, the host transmission message structure is the same differing only by the
read command following the period.

Table 30 Host Transmission

Character #

1

2

3

4

5

6

7

8

9

10

DESC

ASCII

STX

Address Separator

Parm Number

CTR

Type

Inquiry ETX

^B

1-99

,

0-9

0-9

0-9

.

<below>

?

^C

Constant Table Read Type Strings:

TTL

=

Parameter Title Text

MAX

=

Maximum Parameter Value

MIN

=

Minimum Parameter Value

DEF

=

Default Parameter Value

BKU

=

Backup Parameter Value

FLD

=

Parameter Field length

CTB

=

Parameter Control Byte

CTR Field

This field should always be a period designating a Constant Table Read of type to follow for the
indicated parameter.

Type Field

Insert the desired constant table read type string into this field. The control will respond with the
appropriate data for the type of read requested. This is a three character field.