This tab contains routines to initialize, verify, and maintain the analyzer.

Automated Maintenance Options - Provides buttons for commonly used maintenance
routines for the machine. Select from the following:

Calibration/Verification - Select this routine to use with the Calibration/Verification kit.

Performance Verification - Select this routine to use with the Performance Verification

kit.

Fluidics Prep - Select this routine to use with Fluidics Prep.

System Shutdown - Select this routine to perform a pre-defined system shutdown.

The commands for these routines display in the command list below the buttons.

This list is used to view the name of the current command, its well location, the reagent,
status, and information about the command. Only the current commands are listed in the box.

Reagents - Displays the reagents and well locations necessary for the selected routine.

You can select kits for the routines from:

• The Calibration Kit drop down menu.

• The Performance Verification Kit drop down menu.

Report - Opens the Reports tab, with the Performance Verification Reports selection

displayed in the Report: list.

Import Kit - Opens the Import Calibration or Performance Kit dialog box. Select a kit

and click Open to import a kit.

Run - Initiates the selected routine.

xPONENT

®

for FLEXMAP 3D

®