Luminex xPONENT 4.2 for FLEXMAP User Manual

Page 21

Advertising
background image

Serial Number

Firmware Version

XYP Heater Temperature

Sheath Pressure

Calibration/Verification Status

Last F3DCAL1 Calibration

Last F3DeCAL1 Calibration

Last F3DCAL2 Calibration

Last F3DCAL3 Calibration

Last F3DVER1 Verification

Last F3DeVER1 Verification

Last F3DVER2 Verification

Last Fluidics Test

Delta Calibration Temp

System Temperature

DD Temperature

CL1 Temperature

CL2 Temperature

CL3 Temperature

CL1 Voltage

CL2 Voltage

CL3 Voltage

DD Voltage

RP1 Voltage

Number of Syringes

Items in this list relating to calibration and verification (CAL/VER) have one of the following
states:

Passed - Indicates that the process completed successfully.

Failed - Indicates that the process was not completed successfully. Failed items appear in

red.

Not Current - Indicates that verifiers are not current. Verifiers are not current if you have

not calibrated the system since the last time you ran the verifiers.

Not Yet Run - Indicates that this process has not yet been run on the machine.

Copy - Copies the system information to the Windows clipboard. You can then paste it into a
text editor such as Notepad.

Save - Opens the Save As dialog box to specify a file name and location to save the system
information file.

Introduction

7

Advertising