Application settings, Lis settings – Luminex xPONENT 4.2 for FLEXMAP User Manual

Page 124

Advertising
background image

Use this tab to configure system settings such as application settings, LIS settings,
maintenance page options, and external analysis program settings.

Application Settings

These settings change some of the basic ways that xPONENT works, allowing customization
of many of the basic functions.
Enable Virtual Keyboard - Enables the touchscreen virtual keyboard.

Allow the application to be minimized - Enables minimization of the xPONENT software

so that the computer’s desktop can be accessed. If this check box is clear, the application’s
minimize button will not function.

Ignore user permission when attempting to exit software - Enables the user to exit the

xPONENT software, regardless if permission has been granted in the User Setup tab.

Add header when exporting data from grids - Adds a header row when exporting the

data.

Require Fluidics as part of successful verification - Verification will only be marked as

complete if the system attempts and passes the fluidics check. That check involves testing
carryover from one well to the next.

Allow 3D View - Enables the 3D view in the Current Batch tab of the Results page.

LIS Settings

This setting is only for those using xPONENT

®

with a Lab Information System (LIS).

Connect to LIS - Enables connection to the LIS.

Browse - Opens the Browse For Folder dialog box. Select a location for the system to

watch for incoming orders.

xPONENT

®

for FLEXMAP 3D

®

110

Advertising