Luminex xPONENT 4.2 for FLEXMAP User Manual

Page 25

Advertising
background image

NOTE: If using a strip-well plate, ensure that a strip is present in the

selected well location, prior to calibration.

5. Place the plate on the plate holder with A1 in the top left corner. Place the off-plate

reagent block on the plate carrier. Place a well-strip (provided with the Calibration and
the Performance Verification kits) in column S1 of the block.

NOTE: Make sure there is no liquid in the plate or the off plate reagent

block.

6. Click Retract to retract the plate carrier.

7. Type a name for the plate in the Plate Name box, or select a saved plate from the Plate

Name list.

NOTE: If a saved plate is selected the results of the new calibration will

override the previous calibration.

8. In the Plate section, click D6 for the 96-well plate. Click H12 for the 384-well plate.

9. In the Reservoir section, click RB1.

10. In the Strip-Wells section, click SD1.

11. Click Auto Adjust Height. The probe automatically adjusts itself and saves it to the plate

you selected.

NOTE: The probe height is automatically set to 0.49 mm. The probe

automatically adjusts this distance from the bottom of the plate or
calibration disks.

NOTE: When you adjust and save the probe height settings for all three

under a plate name, all areas retain the adjustment.