Local (front leveling panel) operation operations – Anritsu 681XXC User Manual

Page 102

Advertising
background image

At the Configure GPIB menu, press

More

to go to

an Additional Configure GPIB Menu (below).

Press

Pwr Mtr Address

to change the address of the

power meter on the GPIB (the power meter’s default
address is 13). Enter the new address, between 1
and 30, using the cursor control key or the data en-
try key pad and the terminator key

Hz

ns

ADRS

The new GPIB address will appear on the display.

Press

Pwr Mtr Select

to select the power meter

model being used. (Supported power meters are the
Anritsu ML2437A, ML2438A, and ML4803A and
Hewlett-Packard 437B, 438A, and 70100A.)

Press

Previous Menu

to return to the main Config-

ure GPIB Menu display.

At the Configure GPIB menu, press

Previous Menu

to return to the System Configuration Menu display.

Creating a Power-Offset Table
The 681XXC must be in CW frequency mode and
fixed (non-swept) power level mode in order to cre-
ate a power-offset table for user level flatness cor-
rection.

Place the signal generator in CW frequency mode by
pressing the main menu key

CW/SWEEP

SELECT

At the resulting menu display, press

CW

. The

681XXC is now in CW frequency mode.

681XXC OM

3-69

LOCAL (FRONT

LEVELING

PANEL) OPERATION

OPERATIONS